TM   5-4220-231-14&P b. Replace (1) Use an allen wrench to   loosen set screw (2),   and remove handle (1) from  new   valve   (4).    Remove hex   nut (3)  from   valve   (4). Retain    removed parts. (2) Lubricate new   O-ring (9),   M83248/2- 904,   with   silicone compound, (Item   9, Appendix D),  and    install   over   threads 9 of positionable elbow (8). Install 4 elbow   (8)  in  valve   (4). 8 (3) Position valve    (4)  from   back    of  panel, 2 and    install   hex   nut   (3)  on   valve    stem. 3 Align    tube prior   to   securing hex    nut. Install    and   secure handle (1)  on   stem by  tightening set    screw (2). 4 2 4-36