Remove relief    valve
Complete Reentry     Control     Report and   Log

TM-5-4220-231-14P Recompression Chamber 3-Person Model RC-100-2 Serial Numbers 08-A-163 Thru 08-A-172 Manual
Page Navigation
  180    181    182    183    184  185  186    187    188    189    190  
TM  5-4220-231-14&P b. Replace (1) Wrap   teflon   tape    (Item   19,  Appendix 14 D)  over   threads and    install   new    relief valve   (15). 15 (2) Install   new   relief   valve    (15)   on   tee (14) (Tee is assembled to positionable union     and     positionable 1 elbow.) -13 12 (3) Lubricate    new  O-ring   (13),   M83248/2- 904,   with   silicone    compound (Item    9, 9 10 Appendix D), and install on positionable union    (12). (4) Install     housing of     valve (1)    on positionable union    (12). (5) Position     valve (1)   through hole     in panel    so   hex   nut   (9)  can    be   installed on    valve     stem     from    front    of   panel. Install    handle (11)   on   stem    with   set screw    (10). ;-@l 3 1 2 4-40