Remove bolts
nstall    primary HP   filter   to   regulator

TM-5-4220-231-14P Recompression Chamber 3-Person Model RC-100-2 Serial Numbers 08-A-163 Thru 08-A-172 Manual
Page Navigation
  229    230    231    232    233  234  235    236    237    238    239  
TM5-4220-231-14&P (3) Remove the two hex socket bolts from the regulator. Position the regulator under the    support bracket; replace both    bolts    and   torque to   50  ft. lbs. 4 (4). Secure dome loaded regulator (1), 2 with   elbow (2)   and adapter (4),    in location on CRS base, using flat washers (14),   lock  washers (13),   and bolts   (12). 14 13 12 C, 5 6 (5) Lubricate    new   O-ring   (8),   M83248/2- 4 010,    with  silicone compound (Item   9, Appendix D)  and   place   O-ring   (8)  in O-ring groove. Position and    tighten tube (5)  to   adapter (4)   by   turning union   nut   (6)  clockwise. Repeat the same procedures to   secure the   other end of   tube (5)   to   the primary high pressure regulator, if required. (6) Lubricate new   O-ring   (10),   M83248/2- 116,   with  silicone compound (Item  9, Appendix D) and   place   O-ring   (10)  in 2 elbow   (2). Position    and   tighten   tube (7)  to  elbow   (2)   by turning   union   nut (3) clockwise. Tighten elbow (2) securely to   regulator (1). 7 1 4 7 3 10 4-89